All Categories
Media: Air
Output Ranges: 5.8 psi (0.4 bar) vacuum to 5.8 psi (0.4 bar) positive pressure
Pressure Resolution: 0.01 Pa; 0.0001 mbar
Process Connection: M20*1.5 or 1/4 inch NPT
Gauge Connection: M20*1.5 or 1/4 inch NPT
Dwyer LPCP-2 Low Pressure Calibration Pump Low pressure calibration pump with 1/2 NPT connections Pressure resolution: 0.01 Pa; 0.0001 mbar Heat insulation and heat preservation, no negative side effects from temperature changes Portable - Easy to operateThe Model LPCP Low-Pressure Calibration Pump is a low air pressure source with the ability to easily adjust and stabilize. This hand pump possesses a pressure range of +- 5.8 psi; uses air as the media, and can adjust the pressure easily with great stability. The LPCP is made up of quick connectors for fast instrument connect and disconnect. The pump has a heat-insulator between the cover and pressure chamber to lessen the heat effect during the micro-pressure calibration. The adjusting resolution is up to 0.01 Pa (0.0001 mbar). These features make the LPCP ideal for calibrating pressure transducers, precision pressure gauges, and other pressure instruments.