X
Hioki P2000 - DC High Voltage Probe
Hioki P2000 - DC High Voltage Probe
Hioki P2000 - DC High Voltage Probe
Hioki P2000 - DC High Voltage Probe
Hioki P2000 - DC High Voltage Probe
Hioki P2000 - DC High Voltage Probe
Hioki P2000 - DC High Voltage Probe
Hioki P2000 - DC High Voltage Probe

Hioki P2000 - DC High Voltage Probe

Product ID : 49878990
4.9 out of 5 stars


Galleon Product ID 49878990
Shipping Weight 0.73 lbs
I think this is wrong?
Model P2000
Manufacturer Hioki
Shipping Dimension 8.9 x 3.43 x 2.28 inches
I think this is wrong?
-
Save 43%
Before ₱ 20,893
11,815

*Price and Stocks may change without prior notice
*Packaging of actual item may differ from photo shown
  • Electrical items MAY be 110 volts.
  • 7 Day Return Policy
  • All products are genuine and original
  • Cash On Delivery/Cash Upon Pickup Available

Pay with

Hioki P2000 - DC High Voltage Probe Features

  • Simply connect to the corresponding HIOKI clamp meter or digital multimeter to achieve high voltage measurement up to CAT III 2000 V

  • Maintenance and maintenance of solar power equipment, maintenance and maintenance of electrical equipment over 1000 V

  • Compatible measuring devices: Digital multimeter DT4261, AC/DC clamp meter CM4375-50, CM4373-50, CM4371-50, AC clamp meter CM4141-50

  • DT4261 in the image is sold separately


About Hioki P2000 - DC High Voltage Probe

P2000 Specifications Maximum Input Voltage: DC 2,000 V: INPUT H - INPUT L, Maximum Rated Voltage [Maximum Rated Voltage Between ground] 1,000 V (Measurement Category IV) Expected Transient Voltage 12,000 V, 2,000 V (Measurement Category III) Expected Transient Voltage 1,5,000 V Input resistance: 20 MΩ ± 5.0% (INPUT H - INPUT L intervals, with OUTPUT terminal open). Overload Protection: DC 2,200 V/AC 2,200 V 1 minute (INPUT H – INPUT L interval), DC 600 V/AC 600 V 1 minute (OUTPUT H – OUTPUT L intervals). [OUTPUT Terminal] 0.2 inch (4 mm) banana terminal. Operating Temperature Range: -5°F to 145°F (-25°C to 65°C). Operating Humidity Range: -5°F to 104°F (-25°C to 40°C): 80% RH or less (non-condensation), 40°C to 65°C: 40°C 80% RH or lower to 65°C 25% RH (do not condensation), Storage temperature and humidity range: -30°C - 70°C, 90% RH or lower (do not condensation) [Compliance Standards] EN 61010 Accessories: Connection cable L4943 x 1, Strap buckle x 2, Strap x 1, Carrying case C0205 x 1, Instruction manual x 1, Usage Precautions x 1. [Model Name] Combination accuracy specification [DT4261] Range 600.0 V: Combination Accuracy: ±0.5% rdg ±0.2 V, Range 2000 V: Combination Accuracy: ±0.5% rdg ±5 V [CM4141-50, CM4371-50, CM4373-50, CM4375-50) Range 600.0 V: Combination Accuracy: ±1.0% rdg ±0.3 V, Range 2000 V: Combination Accuracy: ±1.0% rdg ±3 V Accessories: [Accessories] Connection cable L4943 x 1, Strap buckle x 2, Strap x 1, Carrying case C0205 x 1, Instruction manual x 1, Usage Precautions x 1.